Manual

Operator’s Manual
922258-00 Rev A 49
Coupling
PRx:COUPLING, PRx:CPL
Command/query
Description The PRx:COUPLING command sets the input coupling for the + and
inputs of the differential amplifier. The valid arguments are AC, Ground and
DC.
The PRx:COUPLING? query returns the input coupling setting of the + and
inputs of the differential amplifier connected to the specified channel.
The input coupling is ignored when PVG Mode is set to V
COMP
. Likewise,
the input coupling argument is not returned from a query when PVG
Mode is set to V
COMP
.
Command Syntax <channel>:CouPLing<+coupling>|,<coupling}|
<channel>:={PR1, PR2, PR3, PR4}
<+coupling>:={AC, DC or GND}
<coupling>:={AC, DC or GND} (ignored when PVG Mode is set to
VCOMP)
Query Syntax <channel>:CouPLing?
Response Format <channel>:CPL<+coupling>,<coupling>
Example The following command sets the + input coupling to DC and the input
coupling to ground for the differential amplifier connected to channel 2.
CMD$=:”PR2:CPL DC,GND”: CALL IBWRT(SCOPE%,CMD$)