Operators Manual User guide
Operator's Manual
CANbus-TD-TDM-OM-E RevB
53
3. Connect the CANbus TD/TDM Oscilloscope
Interface Module (OIM) to the EXT input of
the instrument. Make sure that the top
(OIM labeled side) is facing up.
Note: Make sure that the OIM is at a right angle
to the connector.
4. Connect the 3-pin plug end of the OIM to
the Trigger Module.
Installing Trigger Couplers
UNDERSTANDING WHICH TRIGGER COUPLER IS INSTALLED IN THE TRIGGER MODULE
The Trigger Coupler is a CAN Transceiver. Therefore, the Trigger Coupler in the Trigger Module
must match the transceiver in the CAN circuit you wish to connect to. If it does not match, the
CAN trigger will not function properly.
To identify the coupler(s) that are installed, do the following:
1. Open the CAN Trigger dialog by touching the Trigger descriptor
box (or by pushing the front panel “Bus Analysis” push button
on Vehicle Bus Analyzers, and then selecting the CAN Trigger