46 922261-00 Rev A
I. Attenuator Matching and Final Check
1. Repeat the Performance Verification procedure to ensure compliance with the warranted
specifications.
2. Perform the Attenuator Matching Procedure listed on page 5-8. Apply calibration seals in
accordance with your quality procedures. This concludes the Adjustment Procedure.
Repeat the Performance Verification procedure to complete the calibration of the AP034.
This concludes the Adjustment Procedure. Repeat the Performance Verification procedure to
complete the calibration of the AP034.