Instructions
AP033 Active Differential Probe
38
Procedure
NOTE: This procedure must be performed with the probe set to Manual gain control. See
Operating with Teledyne LeCroy Oscilloscopes on page 13 for more information about
controlling probe settings in different software versions.
A. Adjust Preliminary Probe DC Balance (R36)
1. From the oscilloscope’s front panel, select Channel 1. Set probe Gain Control to
Manual, Attenuation to ÷10, and Gain to X10.
2. If necessary, set the probe offset to 0.000 V.
3. Connect one end of a BNC cable to the probe end of the ProBus Extension Cable.
Attach the precision 50 Ω terminator to the opposite end of the BNC cable.
4. Insert the banana plugs of the precision 50 Ω terminator into the input terminals of
the DMM.
5. Insert the two pin shorting plug into the socket on the Logic board. Use the two holes
closest to the probe cable. (See Figure 8).
Figure 9, AP033 Logic Board
6. To enter the Cal Mode, push any one of the four buttons that protrude
through holes in the Logic board. Entry into Cal Mode is confirmed by the
absence of an audible “click” when you push the button.
7. Adjust DC Bal (R36) on the probe tip (Figure 9) to obtain a DMM reading of 0 V ±10
mV.
NOTE: This voltage may not be stable because of air currents near the probe. The
average voltage should be 0 V ±10 mV.