Instructions
Operator’s Manual
33
AP033 Performance Verification Test Record, Page 2
Serial #:______________________________ Date: _______________
Step
Description
Intermediate Data
Test Result
Gain Accuracy
A-7 Sine Wave Generator Output Voltage ___________ mV
A-13
Probe Output Voltage
___________ mV
A-15
X1 Gain Error (Test limit ≤ ±2%)
__________ %
A-18
Probe Output Voltage
___________ mV
A-20
X10 Gain Error (Test limit
≤
±2%)
__________ %
A-22
Expected Probe Output Voltage
___________ mV
A-24
Measured Probe Output Voltage
___________ mV
A-26
÷10 Gain Error (Test limit ≤ ±2%)
__________ %
High Frequency Common Mode Rejection Ratio (CMRR)
B-11
Probe Output Voltage at 250 MHz
___________ mV
B-12
Differential Mode Gain at 250 MHz
___________ mV
B-21
Common Mode Signal at 250 MHz
___________ mV
B-23
Common Mode Gain at 250 MHz
___________ mV
B-25
CMRR at 250 MHz (Test Limit ≥ 5:1)
__________ :1
Low Frequency Common Mode Rejection Ratio
C-8
Common Mode Signal at 70 Hz
___________ mV
C-9
CMRR at 70 Hz (Test Limit ≥ 3160:1)
__________ :1
C-15
Common Mode Signal at 1 MHz
___________ mV
C-16
CMRR at 1 MHz (Test Limit ≥ 1000:1)
__________ :1