Instructions

Operator’s Manual
15
Adding Offset
The AP033 has true differential offset capability. This allows you to remove a DC bias
voltage from the differential input signal while maintaining DC coupling. By using probe
offset rather than offset at the oscilloscope inut, the full dynamic range of the probe
remains centered around the offset level. Add or remove offset using the oscilloscope
OFFSET controls.
Figure 6, Effect of Differential Offset
NOTE: In any software version later than legacy 8.1.0, the oscilloscope OFFSET
controls change the probe offset, rather than the offset at the oscilloscope input. The
OFFSET buttons on the probe body are disabled, although in Manual mode all other
buttons are functional.
In legacy software versions 8.0.0 and earlier, or when used with the ADPPS Power
Supply on instruments that lack ProBus support, probe offset must be controlled using
the OFFSET buttons on the probe body. The offset can be returned to zero by briefly
pressing the
and
OFFSET buttons at the same time.
AutoZero
The AP033 incorporates an AutoZero function to remove the DC offset from the output.
When power is first applied, the probe will return to the internal values resulting from the
last AutoZero cycle. After several minutes of warm up, or when the probe is exposed to a
large shift in ambient temperature, some DC offset drift may occur from thermal effects
in the amplifier circuitry. To adjust for this, manually invoke AutoZero.
CAUTION: Always remove the probe from the test circuit before initiating
AutoZero.