82 ISSUED: June 2013 923133 Rev A
By enabling cursor
tracking, the difference
cursors are swept
across the hist ogram.
As expected, the space
between the 3 T and 4T
distribution s is the
shortest, be cause of
inter-symbol
interference and the
many 3T widt hs. The
spacing is 138 ns.
To measure the spread of widths for the distributions, set the measure mode to My Measure and
configure the parameters:
1.
P2: average
of F2
2. P3: high of F2
3. P4: low of F2
4.
P5: range of
F2
Becau
se parameter measurements are performed only on those portions of the waveform
between the parameter cursors, activate tracking so that they can be swept across the histogram.
Set the difference between the cursors so that they encompass one clock period. In this case, the
histogram is shown at 0.2 µs per division. Set the difference between the parameter cursors to:
2315 10
210
116
9
6
.
.
.
⋅
⋅
=
−
−
divisions