User's Manual
Appendix C: Performance Verification
CĆ36
AWG2021 User Manual
Press the numeric key 1, and press the units key V to select an
amplitude of 1 V.
c. Repeat substeps 4a through e, checking to the follow limits:
Rise time 4.2 ns, maximum. . . . . . . . . . . . . . . . . . . . . . . .
Aberrations 0.4 div., maximum. . . . . . . . . . . . . . . . . . . . . .
Flatness 0.15 div., maximum. . . . . . . . . . . . . . . . . . . . . . . . .
Fall time 4.2 ns, maximum. . . . . . . . . . . . . . . . . . . . . . . .
6. Check pulse response for CH2 (Option 02): If the AWG2021 has a second
channel, repeat this Pulse Response Check procedure using the AWG2021
CH2 output and selecting the waveform and setting controls for CH2.
7. End procedure: Remove the connections.
SYNC Out and MARKER Out Amplitude Checks
These procedures check the amplitude of the SYNC Out and MARKER Out
signals.
Electrical
Characteristic Checked
$%!& $#$#" & !! !! #$
'(
Equipment Required 'W % ""
Prerequisites #" # !! $"#" "# (
1. Install test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 SYNC Out
connector through the coaxial cable to the CH1 vertical input connector
on the oscilloscope (see Figure C-17).
""
540
TDS
AWG2021
Figure CĆ17: Initial Test Hookup