User's Manual
Appendix C: Performance Verification
AWG2021 User Manual
b. Set frequency counter controls:
CHANNEL A
Termination 50 W. . . . . . . . . . . . . . . . . .
Slope Negative. . . . . . . . . . . . . . . . . . . . . . .
Attenuation X5. . . . . . . . . . . . . . . . . . .
Coupling AC. . . . . . . . . . . . . . . . . . . . .
FREQ A
2. Set AWG2021 controls and select the waveform:
a. Initialize AWG2021 controls: Push UTILITYMiscConfig...Re-
set to FactoryO.K.
b. Select the waveform file:
Push SETUPWaveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to select the CLK_FREQ.WFM file.
Push ENTER.
3. Check clock frequency accuracy:
a. Check clock frequency accuracy at current clock frequency setting:
Check that the frequency counter reading falls between 249.9875 MHz
and 250.0125 MHz.
b. Check clock frequency accuracy for different clock frequency settings:
Select Clock from the bottom of the SETUP menu.
Turn the general purpose knob (or press the numeric and units keys,
and push ENTER) to select the first clock frequency listed in Table
C-3.
Check that the frequency counter reading is within the frequency
range listed in the table for the clock frequency setting.
Repeat this step for each clock frequency and frequency range listed
in Table C-3.