User's Manual
Appendix C: Performance Verification
AWG2021 User Manual
CĆ23
 
FG 5010
8842A
AWG2021
Figure CĆ11: External AM Operation Initial Test Hookup
c. Set DMM controls:
Mode VDC. . . . . . . . . . . . . . . . . . . . . . . . . . . 
Range 20. . . . . . . . . . . . . . . . . . . . . . . . . . 
Inputs Front. . . . . . . . . . . . . . . . . . . . . . . . . . 
d. Set function generator controls:
Function Square. . . . . . . . . . . . . . . . . . . . . . . . 
Mode Continuous. . . . . . . . . . . . . . . . . . . . . . . . . . . 
Parameter
Frequency 1 kHz. . . . . . . . . . . . . . . . . . . . 
Amplitude 0 V. . . . . . . . . . . . . . . . . . . 
Offset 1 V. . . . . . . . . . . . . . . . . . . . . . . 
Output Off. . . . . . . . . . . . . . . . . . . . . . . . . . 
2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITYMiscConfig...Re-
set to FactoryO.K.
b. Modify AWG2021 default settings:
 Push MODEAutostep.
c. Select waveform file:
 Select Select Autostep File from the side menu to choose a
waveform file for CH1.