User's Manual
Appendix C: Performance Verification
AWG2021 User Manual
CĆ21
540
TDS
AWG2021
Figure CĆ10: Autostep Mode Initial Test Hookup
b. Set the oscilloscope controls:
Vertical CH1. . . . . . . . . . . . . . . . . . . . . . . . .
CH1 coupling DC. . . . . . . . . . . . . . . . .
CH1 scale 0.5 V/div.. . . . . . . . . . . . . . . . . . . .
CH1 input impedance 50 W. . . . . . . . . .
Horizontal
Sweep 2 ms/div.. . . . . . . . . . . . . . . . . . . . . . .
Trigger
Source CH1. . . . . . . . . . . . . . . . . . . . . .
Coupling DC. . . . . . . . . . . . . . . . . . . . .
Slope Positive. . . . . . . . . . . . . . . . . . . . . . .
Level 100 mV. . . . . . . . . . . . . . . . . . . . . . .
Mode Normal. . . . . . . . . . . . . . . . . . . . . . .
2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITYMiscConfig...Re-
set to FactoryO.K.
b. Modify AWG2021 default settings:
Push MODEAutostepSlope to highlight Positive.
Select Level from the side menu, and turn the general purpose knob
to select 1 V.
Select Select Autostep File from the side menu to choose from the
file list for CH1.
Turn the general purpose knob to highlight the MODE_AST.AST
file.
Push ENTER.