User's Manual
Appendix C: Performance Verification
CĆ20
AWG2021 User Manual
 Push MODEWaveform AdvanceSlope to highlight Positive.
 Select Level from the side menu, and turn the general purpose knob
to select a 1.0 V level.
 Select Impedance from the side menu to highlight 50 W.
c. Select waveform file:
 Push SETUPWaveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
 Highlight the MODE_ADV.SEQ file using the general purpose
knob.
 Push ENTER to select the file.
3. Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check waveform advance: Repeatedly push the AWG2021 MANUAL
TRIGGER button, and check that the oscilloscope displays a continuous
sine wave that switches between two frequencies at each manual trigger.
5. End procedure: Disconnect the oscilloscope.
Electrical
Characteristic Checked
     #$
Equipment Required ! #W "    
Prerequisites          $
1. Install test hookup and set test equipment controls:
a. Hook up the oscilloscope:
 Connect the AWG2021 CH1 output through a coaxial cable to the
oscilloscope CH1 vertical input.
 Connect the AWG2021 CH1 SYNC Out output through a coaxial
cable to the oscilloscope CH1 vertical input (see Figure C-10).
Check Autostep Mode