User's Manual

Appendix C: Performance Verification
AWG2021 User Manual

2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITYMiscConfig...Re-
set to FactoryO.K.
b. Modify the AWG2021 default settings:
Push MODEGatedPolarity to highlight Positive.
Select Impedance from the side menu to highlight 1 MW.
c. Select the file:
Push SETUPWaveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
Highlight the MODE.WFM file, using the general purpose knob.
Push ENTER to select the file.
3. Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check gated mode with manual trigger: Push and hold the AWG2021
MANUAL TRIGGER button, and check that the oscilloscope continuously
displays a sine wave while the MANUAL TRIGGER button is pushed.
5. Check gated mode with gate signal:
a. Apply gate signal: Turn function generator output on.
b. Check gated mode with positive gate signal: Check that the oscilloscope
displays a sinewave when the gate signal is high.