User's Manual
Appendix C: Performance Verification
AWG2021 User Manual
CĆ15
c. Select the file:
Push SETUPWaveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to highlight the MODE.WFM file.
Push ENTER to select the file.
3. Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check burst count: Push the AWG2021 MANUAL TRIGGER button and
check that when the button is pushed, the oscilloscope displays three cycles
of sine wave.
5. End procedure: Disconnect the oscilloscope.
Electrical
Characteristic Checked
#$
Equipment Required #W " W
!
Prerequisites ! $
1. Install test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector
through the coaxial cable to the CH1 vertical input connector on the
oscilloscope.
b. Hook up the function generator: Connect the function generator output
to both the AWG2021 TRIGGER INPUT and the oscilloscope CH2
input through a coaxial cable, precision termination, and a dual input
coupler (see Figure C-6).
Check Gated Mode