Appendix C: Performance Verification
AWG2021 User Manual
CĆ11
2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITYMiscConfig...Re-
set to FactoryO.K.
b. Select the file:
Push SETUPWaveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to display the list of waveform files
and highlight the MODE.WFM file.
Push ENTER to select the file. This button is located to the
lower-right of the numeric keypad.
3. Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check cont mode: Check that the amplitude of the sine wave displayed on
the oscilloscope is 5 vertical divisions and that 5 cycles of the waveform are
displayed.
5. If Option 02 is installed (Option 02 adds the CH2 output channel): Repeat
this procedure, connecting the oscilloscope to the AWG2021 CH2 output
connector, pushing Waveform Sequence to select the waveform for CH2,
and turning on the CH2 output.
6. End procedure: Disconnect the oscilloscope.
Electrical
Characteristic Checked
$%
Equipment Required " $W # !
Prerequisites ! %
1. Install the test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector
through the coaxial cable to the CH1 vertical input connector on the
oscilloscope.
b. Hook up the function generator:
Connect the AWG2021 TRIGGER INPUT connector through a
coaxial cable to the function generator output connector (see
Figure C-4).
Check Triggered Mode