User's Manual

Appendix C: Performance Verification
CĆ10
AWG2021 User Manual
Operating Mode Checks
These procedures check operation of the Cont, Triggered, Burst, Gated,
Waveform Advance, and Autostep modes.
Electrical
Characteristic Checked
      "#
Equipment Required "W !    
Prerequisites          
#
1. Install the test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector
through the coaxial cable to the CH1 vertical input connector on the
oscilloscope (see Figure C-3).

AWG2021
540
TDS
Figure CĆ3: Cont Mode Initial Test Hookup
b. Set the oscilloscope controls:
Vertical CH1. . . . . . . . . . . . . . . . . . . . . . . . .
CH1 coupling: DC. . . . . . . . . . . . . . . .
CH1 scale 0.2 V/div.. . . . . . . . . . . . . . . . . . . .
CH1 input impedance: 50 W. . . . . . . . . .
Horizontal
Sweep 500 ms/div.. . . . . . . . . . . . . . . . . . . . . . .
Trigger
Source CH1. . . . . . . . . . . . . . . . . . . . . .
Coupling DC. . . . . . . . . . . . . . . . . . . . .
Slope Positive. . . . . . . . . . . . . . . . . . . . . . .
Level +100 mV. . . . . . . . . . . . . . . . . . . . . . .
Mode Auto. . . . . . . . . . . . . . . . . . . . . . .
Check Cont Mode