User manual
Specifications
Table 1: Electrical and mechanical specifications
Characteristic
Bandwidth (–3 dB) DC to 50 MHz
System atten–
uation accuracy
10:1 ±3.2%
Compensation range
15 pF – 25 pF
System input resistance
@DC
10 MΩ ±1.5%
System input capacitance
<12 pF
System rise time (typical)
<7.0 ns
Propagation delay ~6.1 ns
Maximum input voltage
300 V
RMS
CAT II
Cable length
1.3 m
Table 2: Environmental specifications
Characteristics Description
Temperature
Operating
Nonoperating
–10°Cto+55°C(14°Fto+131°F)
–51 °C to +71 °C (–60 °F to +160 °F)
Humidity
Operating and
Non-Operating
5% to 95% relative humidity (%RH) up
to +30 °C (86 °F), 5% to 65% RH above
+30° C up to +55 °C (131 °F)
Altitude
Operating
Nonoperating
3.0 km (10,000 ft) maximum
12.2 km (40,000 ft) maximum
Performance Graphs
Table 3: Certifications and compliances
Characteristics Description
EC
Declaration
of
Conformity
Compliance was demonstrated to the following
specification as listed in the Official Journal of
the European Communities:
Low Voltage Directive 2006/95/EC:
EN61010-031: 2002
Safety
Standards
UL61010-031;2007 (AM1)
CAN/CSA C22.2 No. 61010-031-07 (AM1)
IEC61010-031; IEC 61010-031/A1:2008
Category Examples of Products in this
Category
CAT III Distribution-level mains, fixed
installation
CAT II
Local-level mains, appliances,
portable equipment
Measurement
Category
Descriptions
CAT I Circuits not directly connected to
mains.
Pollution
Degree 2
Do not operate in environments where cond–
uctive pollutants may be present (as defined in
IEC 61010-1). Rated for indoor use only.
Equipment Recycling.
This product complies
with the European Union’s requirements
according to Directive 2002/96/EC on waste
electrical and electronic equipment (WEEE). For
more information about recycling options, check
the Support/Service section of the Tektronix
Web site (www.tektronix.com).
Safety Summary
Review the following safety precautions to avoid injury and
prevent damage to this product or any products connected
to it. To avoid potential hazards, use this product only as
specified. Using the probe or accessories in a manner not
specified could result in a shock or fire hazard.
To Avoid Fire or Personal Injury
Ground-Referenced Oscilloscope Use.
Do not fl oat the
reference lead of this probe when using with ground
referenced oscilloscopes (for example, TBS, DPO, MSO,
and TDS series oscilloscopes). The reference lead must be
connected to earth potential (0 V).
TPS2000 Series Oscilloscope Use.
Do not float the reference
lead of this probe above the rated float voltage (30 V
RMS
).
Connect and Disconnect Properly.
Connect the probe output
to the measurement instrument before connecting the probe
to the circuit under test. Disconnect the probe input and
the probe reference lead from the circuit under test before
disconnecting the probe from the measurement instrument.
Avoid Electric Shock.
Do not connect or disconnect probes
or test leads while they are connected to a voltage source.
Observe All Terminal Ratings.
To avoid fire or shock hazard,
observe all ratings and markings on the product. Consult
the product manual for further ratings information before
making connections to the product.
Avoid Electric Shock.
When using probe accessories, never
exceed the lowest rating of the probe or its accessory,
whichever is less, including the measurement category and
voltage rating.
Inspect the Probe and Accessories.
Before each use, inspect
the probe and accessories for damage (cuts, tears, defects
in the probe body, accessories, cable jacket, etc.). Do not
use if damaged.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Safety Terms and Symbols Terms in This Manual.
These terms may appear in this manual:
WARNING.
Warning statements i dentify conditions or
practices that could result in injury or loss of life.
CAUTION.
Caution statements identify conditions or
practices that could result in damage to this product or
other property.
Symbols on the Product.
These symbols may appear on
the product:
Contacting Tektronix
Web site: www.tektronix.com
Phone: 1-800-833-9200
Address: Tektronix, Inc.
Department or name (if known)
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
Email:
techsupport@tektronix.com
Warranty Information
For warranty information, go to
www.tektronix.com/warranty
Copyright © Tektronix, Inc. All rights reserved. www.tektronix.com