Datasheet
Specifications
All specifications apply to all models unless noted otherwise.
TDS3012C TDS3014C TDS3032C TDS3034C TDS3052C TDS3054C
Input channels 2 4 2 4 2 4
Bandwidth 100 MHz 100 MHz 300 MHz 300 MHz 500 MHz 500 MHz
Rise time (typical) 3.5 ns 3.5 ns 1.2 ns 1.2 ns 0.7 ns 0.7 ns
Sample rate on each channel 1.25 GS/s 1.25 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s 5 GS/s
Record length 10 kpoints
Vertical system
Hardware bandwidth limits
TDS3012C
TDS3014C TDS3032C TDS3034C TDS3052C TDS3054C
20 MHz 20 MHz 20 MHz, 150 MHz 20 MHz, 150 MHz 20 MHz, 150 MHz 20 MHz, 150 MHz
Input coupling AC, DC, GND
Input impedance 1 MΩ in parallel with 13 pF or 50 Ω
Input sensitivity range
1 MΩ 1 mV/div to 10 V/div
50 Ω 1 mV/div to 1 V/div
Vertical resolution 9 bits
Maximum input voltage
1 MΩ 150 V
RMS
with peaks at ≤400 V
50 Ω 5 V
RMS
with peaks at ≤30 V
DC gain accuracy ±2%
Position range ±5 div
Horizontal system
Seconds/division range
TDS3012C
TDS3014C TDS3032C TDS3034C TDS3052C TDS3054C
4 ns to 10 s 4 ns to 10 s 2 ns to 10 s 2 ns to 10 s 1 ns to 10 s 1 ns to 10 s
Time base accuracy ±20 ppm over any 1 ms time interval
TDS3000C Series Digital Phosphor Oscilloscopes
www.tektronix.com 5