User manual

Reference
Options Settings Comments
BW Limit
20 MHz
1
,Off
Limits the ban
dwidth to reduce display noise;
lters the signal to reduce noise and other
unwanted high frequency components
Volts/Div Coarse, Fine Selects the resolution of the Scale (Volts/Div)
knob
Coarse denes a 1-2-5 sequence. Fine
changes the resolution to small steps
between the
coarse settings
Probe
See the next
table
Push to adjust Probe options
Invert
On, Off Inverts (
ips) the w aveform with respect to
the reference level
1
Effective bandwidth is 6 MHz with a P2220 probe set to 1X.
The option for volta
ge and current probes is different: Attenuation or Scale.
Probe options Setting s Comments
Probe Voltage
Attenuation
1X, 10X, 20X , 50X, 100X,
500X, 1000X
Set to match the attenuation
factor of the voltage probe
to ensure correct vertical
readouts
Probe Current Scale 5 V/A, 1 V/A, 500 mV/A,
200 mV/A, 100 mV/A,
20 mV/A, 10 mV/A, 1 mV /A
Set to match the scale of
the current probe to ensure
correct v ertical readouts
Back Returns to the previous
menu
Knobs
Vertical Position Knobs. Use the Vertical Position knobs to move the channel
waveformsupordownonthescreen.
Vertical Scale (volts/division) Knobs. Use the Vertical Scale knobs to control how
the o scilloscope amplies or attenuates the source signal of channel waveforms.
WhenyouturnaVerti ca l S cale knob, the oscilloscope increases or decreases the
vertical size of the waveform on the screen.
Vertical Measurement Overrange (Clipping). Waveforms that extend beyond the
screen (overrange) and display a ? in the measurement readout indicates an
invalid value. Adjust the vertical scaling to ensure the readout is valid.
Key Points
Ground Coupling. Use Ground coupling to display a zero-volt waveform.
Internally, the channel input is connected to a zero-volt reference level.
Fine Resolution. The vertical scale readout displays the actual volts/division
setting while in the ne resolution setting. Changing the setting to coarse does not
change the vertical scale until the Vertical Scale control is adjusted.
TDS2000C and TDS1000C-EDU Series Oscilloscope User Manual 105