Settings to avo
id aliasing in Sample mode
Time base Samples per second Maximum
2.5 ns
2GS/s
200.0 MHz
†
5.0 to 250.0 n
s
1GS/sor2GS/s
*
200.0 MHz
†
500.0 ns
500.0 MS/s
200.0 MHz
†
1.0 μs
250.0 MS/s
125.0 MHz
†
2.5 μs
100.0 MS/s
50.0 MHz
†
5.0 μs
50.0 MS/s
25.0 MHz
†
10.0 μs
25.0 MS/s
12.5 MHz
†
25.0 μs
10.0 MS/s
5.0 M Hz
50.0 μs
5.0 MS/s
2.5 M Hz
100.0 μ
s
2.5 MS/s
1.25 MH
z
250.0 μ
s
1.0 MS/s
500.0 k
Hz
500.0
μs
500.0 kS/s
250.0
kHz
1.0 m
s
250.0 kS/s
125.
0kHz
2.5 m
s
100.0 kS/s
50.0
kHz
5.0
ms
50.0 kS/s
25.
0kHz
10.
0ms
25.0 kS/s
12.
5kHz
25
.0 m s
10.0 kS/s
5.
0kHz
50
.0 m s
5.0 kS/s
2.
5kHz
1
00.0 ms
2.5 kS/s
1
.25 kHz
2
50.0 ms
1.0 kS/s
5
00.0 Hz
500.0 ms
500.0 S/s
250.0 Hz
1.0 s
250.0 S/s
125.0 Hz
2.5 s
100.0 S/s
50.0 Hz
5.0 s
50.0 S/s
25.0 Hz
10.0 s
25.0 S/s
12.5 Hz
25.0 s
10.0 S/s
5.0 Hz
50.0 s
5.0 S/s
2.5 Hz
*
Depending on the oscilloscope model.
†
Bandwidthreducedto6MHzwithaP2220probesetto1X.
Taking Measurem ents
The oscilloscope displays graphs of voltage versus time and can help you to
measure the displayed waveform.
There are several ways to take measurements. You can use the graticule, the
cursors, or an automated measurement.
TDS2000C and TDS1000C-EDU Series Oscilloscope User Manual 25