Perform this functional check to verify that your oscilloscope is operating
correctly.
ON/OFF button
1. Power on the o scilloscope.
Push the Default Setup button.
The default
Probe option attenuation setting is
10X.
Default Setup button
PROBE COMP
2. Connect the TPP0101/TP0201 probe to channel
1ontheo
scilloscope. To do this, align the slot
in the probe connector with the key on the CH 1
BNC, push to connect, and twist to the right to
lock th
e probe in place.
Connect the probe tip and reference lead to the
PROBE COMP terminals.
3. Push t
he AutoSet button. Within a few seconds,
you should see a square wave in the display of
about 5V peak-to-peak at 1 kHz.
Push t
he 1 channel 1 m enu button on the front
panel twice to remove channel 1, push the 2
channel 2 m enu button to display channel 2, and
repe
at steps 2 and 3. For 4-channel models,
repeat for 3 and 4.
4 TDS2000C and TDS1000C-EDU Series Oscilloscope User Manual