User Manual
Appendix A: Specifications
Table 3: Vertical Specifications (cont.)
Characteris
tic
Description
V/div values are accurate for probe attenuation settings of 1X. No probe should be installed for these measurements.
TDS2001C TDS2002C, 2004C TDS1012C-EDU,
TDS2012C, 2014C
TDS2022C, 2024C
DC to >50 MHz for 5 mV/div
through 5 V/div settings
with bandw
idth limi t at full.
<5 mV/div settings are limited
to 20 MHz BW
DC to >70 MHz for 5 mV/div
through 5 V/div settings
with bandw
idth limit at full.
<5 mV/div settings are
limitedto20MHzBW
DC to >100 MHz for 5 mV/div
through 5 V/div settings
with bandw
idth limit at full.
<5 mV/div settings are limited
to 20 MHz BW
DC to >200 MHz for 5 mV/div
through 5 V/div settings with
bandwidth
limit at full with
temperature between 0 and
35 °C.
DC to >160
MHz from
5 mV/div through 5 V/div
settings with bandwidth
limit at
full for temperatures
between 0 and 50 °C. <
5mV/div settings are limited
to 20 MH
zBW
TDS1001C-EDU TDS1002C-EDU
Analog Bandwidth, DC
Coupled, Sa
mple or
Average
DC to >40 MHz for 5 mV/div
through 5 V/div settings
with b
andwidth limit at full.
<5 mV/div settings are limited
to 20 M Hz BW.
DC to >60 MHz for 5 mV/div
through 5 V/div settings
with b
andwidth limit at full.
<5 mV/div settings are
limitedto20MHzBW
The Analog Bandwidth when the instrument is DC coupled. V/div values are accurate for probe attenuation settings of 1X. No
probe should be installed for these measurements.
TDS1001C-EDU, TDS2001C TDS1002C-EDU,
TDS2002C, 2004C
TDS1012C-EDU, TDS2012C, 2014C, 2022C, 2024C
Analog Bandwidth, DC
Coupled, Peak Detect
DC to >30 MHz for 5 mV/div
through 5 V/div settings
wit
h bandwidth limit at full.
Setting less than 5 mV/div
are limited to 20 MHz BW
DC to >50 MHz for 5 mV/div
through 5 V/div settings
wit
h bandwidth limit at full.
Setting less than 5 mV/div
are limited to 20 MHz BW
DC to >75 MHz for 5 mV/div through 5 V/div s ettings with
bandwidth limit at full. Setting less than 5 m V/div are limited
to 2
0MHzBW
An
alog Bandwidth
Selectio ns
20 MHz BW Limit ON/OFF
Lo
wer Frequency Limit, AC
Coupled
≤ 10 Hz
≤
1 Hz when 10X passive probes are used.
Rise time is generally calculated from the following formula: Rise time in ns = 350 / Bandwidth in MHz
TDS2001C TDS2002C, 2004C TDS1012C-EDU,
TDS2012C, 2014C
TDS2022C, 2024C
= 7.0 ns = 5.0 ns = 3.5 ns = 2.1 ns
TDS1001C-EDU TDS1002C-EDU
Rise Time, typical
=8.8ns =5.8ns
The capability of the instrument to capture single event pulses using Peak Detect Acquisition Mode.
The minimum single pulse widths for guaranteed 50% or greater amplitude capture are as follows:
Sec/Div Setting
Minimum Pulse Width
50 s/div to 5 us/div
TDS1002C-EDU, TDS1012C-EDU, TDS2002C, TDS2004C,
TDS2012C, TDS2014C, TDS2022C, TDS2024C, TDS2001C
12 ns
Peak Detect Mode Pulse
Response
TDS1001C-EDU, TDS2001C
13 ns
DC Gain Accuracy, Sample
or Average Acquisition
±3%, 5 V/div through 10 mV/div.
±4%, 5 mV/div and 2 mV/div.
TDS2000C and TDS1000C-EDU Series Oscilloscope User Manual 109