User Manual

Command Descriptions
Related Commands
DISplay:INVer
t,
Arguments
<NR1> is an integer in the range from 1 through 100. The larger the value, the
greater the screen contrast.
Examples
DISPLAY:CONTRAST 63 sets the display contrast to 63%.
DISplay:FORMat
Sets or queries the oscilloscope display format. This command is equivalent to
setting the Format option in the Display menu.
Group
Display
Syntax
DISplay:FORMat { XY | YT }
DISplay:FORMat?
Arguments
XY displays the voltage of channel 1 (horizontal axis) against the voltage of
channel 2 (vertical axis).
NOTE. Setting the display format to XY turns cursors off. Sending the
CURSor:FUNCtion command when the display format is XY causes the
oscilloscope to generate event 221 (Settings conict) and leaves the display in
XY format.
YT sets the display to vertical values versus time format and is the default mode.
Examples
DISPLAY:FORMAT YT selects a voltage versus time format for the display.
DISPLAY:FORMAT? m ight return XY for the display format.
DISplay:INVert
Sets or queries the TDS1000, and TDS1000B display format. This command is
equivalent to setting the Display Style option in the Utility Options menu for
the TDS1000, and TDS1000B Series.
The TDS2000, TDS2000B, TDS2000C, and TPS2000 Series do not have a
Display Style option, but accept the DISplay:INVert command and query for
compatibility. The query always returns OFF.
2-70 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer