User Manual

Command Descriptions
DATa:WIDth
Sets the number of bytes per waveform d ata point to be transferred when
executing the CURVe command. (Changing DATa:WIDth may change the
following WFMPre parameters: BIT_Nr, BYT_Nr, YMULt, YOFf, and YZEro.)
Group
Waveform
Syntax
DATa:WIDth <NR1>
DATa:WIDth?
Related Commands
CURVe, WFMPre:BIT_Nr,
Arguments
<NR1> = 1 sets the number of bytes per waveform data point to 1 byte (8 bits).
<NR1> = 2 sets the number of bytes per waveform data point to 2 bytes (16 bits).
If DATa:WIDth is set to 2, the least signicant byte is always zero.
Examples
DATA:WIDTH 1 sets the data width to 1 byte per data point for CURVe data.
DATALOGging? (Query Only)
Returns all the data logging parameters.
Group
Data Logging
Syntax
DATALOGging?
DATALOGging:DURAtion
Sets the duration of the data logging, in minutes.
Group
Data Logging
Syntax
DATALOGging:DURAtion 30, 60, 90, 120, 150, 180, 210, 240,
270, 30 0, 330, 360, 390, 420, 450, 480
DATALOGging:DURAtion?
Related Commands
DATALOGging:SOURCE, DATALOGging:STATE
2-64 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer