DATA:START 1 0 s
pecifies that the waveform transfer will begin with data point 10.
DATA:START? might return 214 as the first waveform d ata point that will be
transferred.
DATa:STOP
Sets or queries the last data point in the waveform that will be transferred when
executing the CURVe? command. This lets you transfer partial waveforms from
the oscilloscope.
When using the CURVe command, the oscilloscope stops reading data when there
is no more data to read or when the 2500 data point limit is reached.
Group
Waveform
Syntax
DATa:STOP <NR1>
DATa:STOP?
Related Commands
CURVe
Arguments
<NR1> is an integer value that ranges from 1 to 2500, and specifies the last data
point that will be transferred. When DATa:STOP is l ess than DATa:STARt, the
values are swapped internally for CURVe.
If you always want to transfer comple te waveforms, set DATa:STARt to 1 and
DATa:STOP to 2500.
Examples
DATA:STOP 150 specifies that the waveform transfer will stop at data point 150.
DATA:STOP? might return 285 as the last data point that will be transferred.
DATa:TARget
Sets or queries the location for storing waveform data transferred from an
external device to the oscilloscope when executing the CURVe command. This
command is equivalent to the DATa:DESTination command and is included here
for compatibility with older Tektronix oscilloscopes.
Group
Waveform
Syntax
DATa:TARget
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-63