User Manual

Command Descriptions
Examples
DATA INIT reini
tializes the waveform data settings to their factory defaults:
DESTINATION = REFA (=TARGET) ENCDG = RIBINARY SOUrce = CH1
START=1STOP=
2500 WIDTH = 1
DATA? might return the following string: DATA:ENCDG
RPBINARY;DESTINATION REFA; SOURCE REFB;S TART 1;STOP
500;WIDTH 2
DATa:DESTination
Sets or que
ries the reference memory location for storing oscilloscope waveform
data that is transferred into the oscilloscope by the CURVe command. This
command is identical to the DATa:TARget command.
Group
Waveform
Syntax
DATa:DESTination REF<x>
DATa:DESTination?
Related Commands
CURVe
Arguments
REF<x> is the reference memory location where the waveform will be stored.
Examples
DATA:DESTINATION REFA stores incoming waveform data into reference
memory REFA.
DATA:DESTINATION? might return REFB as the waveform location that is
currently selected.
DATa:ENCdg
Sets or queries the format of the waveform data. This command is equivalent
to setting W FMPre:ENCdg, WFMPre:BN_Fmt, and WFMPre:BYT_Or. (See
Table 2-31.)
Setting the DATa:ENCdg value causes the corresponding WFMPre values to
update. Setting the WFMPre value causes the corresponding DATa:ENCdg
values to update.
Group
Waveform
2-60 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer