User Manual
Command Descriptions
Conditions
This command ap
plies to the TDS1000B, TDS2000B, TDS2000C, and TPS2000
Series only.
Group
Cursor
Syntax
CURSor:VBArs:HPOS<x>?
Related Commands
CURSor:HBArs:UNIts?
Returns
<NR3> indicates the amplitude value at the selected position.
Examples
CURSOR:VBARS:HPOS1? might return 1.37, indicating the value of one vertical
bar tic.
CURSor:VBArs:POSITION<x>
Positions a vertical bar cursor. The unit is specified by the CURSor:VBArs:UNIts
command, and can be in units of seconds or frequency (Hertz). If the cursor
source is an FFT math waveform, CURSor:VBArs:POSITION is always in Hertz,
regardless of the value set by CURSor:V BArs:UNIts.
NOTE. If Trigger View is active, the query form returns 9.9E37 and generates
event 221 (Settings conflict).
Group
Cursor
Syntax
CURSor:VBArs:POSITION<x> <NR3>
CURSor:VBArs:POSITION<x>?
Arguments
<x> specifies which cursor to position. Correct values are 1 and 2.
<NR3> specifies the cursor position in the units specified by the
CURSor:VBArs:UNIts command. The position is relative to the trigger except
when the cursor source is a math FFT waveform. The cursor position is limited to
the graticule whenever an attempt is made to move it outside the graticule.
Examples
CURSOR:VBARS:POSITION2 9.00E-6 positions the second vertical bar cursor
at 9ms.
2-56 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer