User Manual

Command Descriptions
Group
Vertical
Syntax
CH<x>:PRObe{1|10|20|50|100|500|1000}
CH<x>:PRObe?
Arguments
1 sets the specied oscilloscope channel to 1X attenuation.
10 sets the specied oscilloscope channel to 10X attenuation.
20 sets the specied oscilloscope channel to 20X attenuation. (TDS1000B,
TDS2000B, TDS2000C, and TPS2000 Series o nly)
50 sets the specied oscilloscope channel to 50X attenuation. (TDS1000B,
TDS2000B, TDS2000C, and TPS2000 Series o nly)
100 sets the specied oscilloscope channel to 100X attenuation.
500 sets the specied oscilloscope channel to 500X attenuation. (TDS1000B,
TDS2000B, TDS2000C, and TPS2000 Series o nly)
1000 sets the specied oscilloscope channel to 1000X attenuation.
Returns
<NR1>
Examples
CH2:PROBE 1000 sets channel 2 to 1000X attenuation.
CH1:PROBE? might return CH1:P ROBE 10.
CH<x>:SCAle
Sets or queries the vertical gain of the specied oscilloscope channel. The value
of <x> can vary from 1 through 4 for 4-channel instruments or 1 through 2 for
2-channel instruments.
This command is equivalent to adjusting the front-panel VOLTS/DIV knob.
Group
Vertical
Syntax
CH<x>:SCAle <NR3>
CH<x>:SCAle?
Related Commands
CH<x>:VOLts
2-48 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer