User Manual

Command Descriptions
CH<x>:POSiti
on
Sets or queries the vertical position of the specied oscilloscope channel. The
value of <x> can vary from 1 through 4 for 4-channel instruments or 1 through
2 for 2-chann
el instruments.
The position voltage value is applied to the signal before digitization. This
command is e
quivalent to adjusting the front-panel VERTICAL POSITION knob.
Group
Vertical
Syntax
CH<x>:POSition <NR3>
CH<x>:POSition?
Arguments
<NR3> is the position in divisions from the center graticule.
Table 2-30: Vertical position ranges using a 1X probe
Vertical position Range
2 mV/div
±1000 divs
5 mV/div
±400 divs
10 mV/div
±200 divs
20 mV/div
±100 divs
50 mV/div
±40 divs
100 mV/div
±20 divs
200 mV/div
±10 divs
500 mV/div
±100 divs
1V/div
±50 divs
2V/div
±25 divs
5V/div
±10 divs
Examples
CH2:POSITION 1.32E0 positions the channel 2 input signal 1.32 divisions
above the center of the display.
CH1:POSITION? might return -1.32E0, indicating that the current position of
channel 1 is at -1.32 divisions.
CH<x>:PRObe
Sets or queries the attenuation factor of the specied channel or voltage probes.
The value of <x> can vary from 1 through 4 for 4-channel instruments or 1
through 2 for 2-channel instruments.
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-47