User Manual
Command Descriptions
This command is
equivalent to setting the Coupling option in the Vertical menu.
Group
Vertical
Syntax
CH<x>:COUPling { AC | DC | GND }
CH<x>:COUPling?
Arguments
AC sets the specified oscilloscope channel to A C coupling.
DC sets the specified oscilloscope channel to D C coupling.
GND sets the specified oscilloscope channel to ground. Only a flat ground-level
waveform is displayed.
Examples
CH1:COUPLING AC establishes AC coupling on channel 1.
CH2:COUPLING? might return DC. Indicating that channel 2 is set to DC coupling.
CH<x>:CURRENTPRObe
Sets or queries the scale factor of the specified channel for c urrent probes. The
value of <x> can vary from 1 through 4 for 4-channel instruments or 1 through
2 for 2-channel instruments.
You can issue this command when CH<x>:YUNit <QString> is set to V. However,
this command only affects the readouts when CH<x>:YUNit <QString> is set
to A.
Conditions
This command applies to the TDS1000B, TDS2000B, TDS2000C, and TPS2000
Series only.
Group
Vertical
Syntax
CH<x>:CURRENTPRObe {0.2|1|2|5|10|50|100|1000}
CH<x>:CURRENTPRObe?
Arguments
0.2 sets the specified oscilloscope channel to 0.2X attenuation.
1 sets the specified oscilloscope channel to 1X attenuation.
2 sets the specified oscilloscope channel to 2X attenuation.
5 sets the sp
ecified oscilloscope channel to 5X attenuation.
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-45