User Manual
Command Descriptions
Returns
Oscilloscope v
ertical settings.
Examples
CH1? might return the following string for channel 1: CH1:SCALE
1.0E0;POSIT
ION 0.0E 0; COUPLING DC;BANDWIDTH OFF;PROBE 1.0E0
CH<x>:BANdwidth
Sets or queries the bandwidth setting of the specified oscilloscope channel. The
value of <x> can vary from 1 through 4 for 4-channel instruments or 1 through
2 for 2-channel instruments.
This command is equivalent to setting the BW Limit option in the Vertical menu.
Group
Vertica
l
Syntax
CH<x>:BANdwidth { ON | OFF }
CH<x>:BANdwidth?
Arguments
ON sets the c hannel bandwidth to 20 MHz.
OFF sets the channel bandwidth to the full bandwidth of the osci
lloscope.
In most acquisition modes, full bandwidth is 60 MHz, 100 MHz, or 200 MHz
(depending on the oscilloscope model). There are exceptions:
For TDS1000, TDS2000, TDS1000B, TDS2000B, TDS2000C, and TPS2000
Series: At vertical scales from 2.00 to 4.99 mV/div (sensitivity at the BNC;
that is, after the probe factor is removed), the full bandwidth is 20 MHz.
For TDS200 Series: At vertical scales of 5 mV/div or less (sensitivity at the
BNC; that is, after the probe factor is removed), the full bandwidth is 20 MHz.
When the acquisition mode is Peak Detect, and the v ertical scale at the BNC
is 10 mV/div or less, the full bandwidth is also 20 MHz.
Examples
CH2:BANDWIDTH ON sets the bandwidth of channel 2 to 20 MHz.
CH1:BANDWIDTH? might return OFF. This indicates there is n o bandwidth
limiting on channel 1.
CH<x>:COUPling
Sets or queries the input attenuator coupling setting of the specified oscilloscope
channel. The value of <x> can vary from 1 through 4 for 4-channel instruments or
1 through 2 for 2-channel instruments.
2-44 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer