*CAL?
Related Commands
CALibrate:INTERNAL
Returns
0 indicates that the self-calibration completed without any errors detected.
Any value other than zero indicates that the self-calibration did not complete
successfully or completed with errors.
Examples
*CAL? performs a self-calibration and might return 0 to indicate that it completed
successfully.
CALibrate:ABOrt (No Query Form)
NOTE. You should only use this command in a qualified service environment. For
more information about the factory calibration sequenc
e, refer to the service
manual for your oscilloscope.
Aborts the factory calibration process. When you abort
the factory calibration, the
oscilloscope restores the ca libration settings to the previous factory calibration
constants stored in nonvolatile memory.
Group
Calibration and Diagnostic
Syntax
CALibrate:ABOrt
Examples
CALIBRATE:ABORTstops the in-process factory calibration procedure.
CALibrate:CONTINUE (No Query Form)
NOTE. You should only use this command in a qualified service environment. For
more information about the factory calibration sequence, refer to the service
manual for your oscilloscope.
Performs the next step in the factory calibration operation.
Group
Calibration and Diagnostic