Command Descriptions
NONE if the AUTO
SET menu is not displayed.
AUTOSet:VIEW (No Query Form)
If the current menu is not the Autoset menu, or if the view is not valid for the
detected waveform, the set command causes the oscilloscope to generate error
221 (Settings conflict).
Conditions
This command applies to the TDS1000, TDS2000, TDS1000B, TDS2000B,
TDS2000C, and TPS2000 Series only.
Group
Miscellaneous
Syntax
AUTOSet:VIEW { MULTICYcle | SINGLECYcle | FFT | RISINGedge |
FALLINGedge | FIELD | ODD | EVEN | LINE | LINENum | DCLIne |
DEFault | NONE }
Arguments
MULTICYcle displays a sine or square wave of several cycles. Default for
sine-like and square-like signals.
SINGLECYcle displays a sine or square wave of approximately one cycle.
FFT displays the FFT of a sine wave.
RISING displays the rising edge of a square wave.
FALLING displays the falling edge of a square wave.
FIELD displays a video signal synchronized on all fields. This is the default for
video signals.
ODD displays a video signal synchronized on odd fields.
EVEN displays a video signal synchronized on even fields.
LINE displays a video signal synchronized on all lines.
LINENum displays a video signal synchronized on the specified line number.
DCLIne returns a query response when the oscilloscope finds a DC level.
DEFault returns a query response when the oscilloscope cannot determine the
signal type.
NONE returns a query response when the AUTOSET menu is not displayed. Set is
ignored.
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-39