User Manual

Command Descriptions
For a detailed d
escription of the Autoset function, refer to the user manual for
your oscilloscope.
Group
Miscellaneo
us
Syntax
AUTOSet EXECute
Arguments
EXECute invokes Autoset.
AUTOSet:ENABLE
Allows educators to disable or enable the Autorange and Autoset functions. The
function can be manually set from the Service Diag menu. To access the menu,
refer to the TDS1000B and TDS2000B Series or to the TDS2000C Series service
manual.
Conditions
This command applies to the TDS1000B, TDS2000B, and TDS2000C Series only.
Group
Miscellaneous
Syntax
AUTOSet:ENABLE
AUTOSet:SIGNAL? (Query Only)
Returns the type of signal discovered by the most recent execution of Autoset.
Group
Miscellaneous
Syntax
AUTOSet:SIGNAL?
Returns
LEVEL if the oscilloscope discovered a DC level.
SINE if the oscilloscope discovered a sine-like waveform.
SQUARE if the oscilloscope discovered a square-like waveform.
VIDPAL if the oscilloscope discovered a PA
L or SECAM standard video signal.
VIDNTSC if the oscilloscope discovered an NTSC standard video signal.
OTHER if the oscilloscope was unable to classify the signal.
2-38 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer