Command Descriptions
ACQuire:NUMA
Cq? (Query Only)
Indicates the number of acquisitions that have taken place since starting
oscilloscope a cquisition. The maximum number of acquisitions that can b e
counted is 2
31
-1. This value is reset to zero when you change most Acquisition,
Horizontal, Vertical, or Trigger arguments that affect the waveform except for
the following:
Changing the trigger level or trigger holdoff when in Sample or Peak Detect
mode does not reset the value
NOTE. Any change made when in Average mode aborts the acquisition and resets
ACQuire:NUMACq to zero.
TDS200: changing the vertical position does not reset the value
TDS1000, TDS2000, TDS1000B, TDS2000B, TDS2000C, and
TPS2000 Series: if the Trigger mode is set to Auto, and the Horizontal Scale
is 10 ms/div or slower, changing the vertical position does not reset the value
NOTE. In Scan mode,
ACQuire:NUMACq? always returns zero.
Group
Acquisition
Syntax
ACQuire:NUMACq?
Returns
<NR1>
Examples
ACQUIRE:NUMACQ? might return 350, indicating that 350 acquisitions took place
since an ACQuire:STATE RUN command was executed.
ACQuire:NUMAVg
Sets the number of oscilloscope waveform acquisitions that make up an averaged
waveform. This command is equivalent to setting the Av erages option in the
Acquire menu.
Group
Acquisition
Syntax
ACQuire:NUMAVg <NR1>
ACQuire:NUMAVg?
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-33