User Manual
Command Descriptions
Returns
The query retur
ns the following:
Volts for volts
U for unknown units (divisions)
dB for decibels
? for unknown mask waveforms units
For the TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series, the query can
return the following additional values:
A for amperes
VA for volt×amperes
AA for amperes×amperes
VV for volts×volts
WFMPre:<wfm>:YUNit
Same as WFMPre:YUNit, except that <wfm> specifies the source/destination
waveform instead of DATa:SOUrce and DATa:DESTINATION.
For set commands, if <wfm> is not a reference waveform, the oscilloscope
generates er
ror 2241.
WFMPre:YZEro
For all model and firmware combinations except the TDS210 or TDS220
oscilloscope (firmware below V 2.00) with a TDS2CMA communications
module, YZEro is a value, expressed in YUNits, used to convert waveform record
values to YUNit values using the following formula (where dl is digitizer levels):
value_in_YUNits = ((curve_in_dl - YOFF_in_dl) * YMUlt) + YZERO_in_YUNits
YZEroisusedwhencalculatingcursorreadouts.
The set form of this command affects the reference waveform specified by
DATA:DESTination.
The query form of this command returns a value for the waveform specified by
DATA:SOUrce. If that waveform is not active or displayed, the query fails and
the oscilloscope generates an execution error with event code 2244 (waveform
requested is not active).
NOTE. For the TDS210 or TDS220 oscilloscope (firmware below V 2.00) with a
TDS2CMA communications module, the set form is ignored. The query always
returns zero.
2-194 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer