User Manual
Command Descriptions
Arguments
<NR3> is the ver
tical scale factor, in Y UNits (usually volts) per sample value.
Returns
A query result of zero has special meaning. A zero indicates the waveform has
unknown vertical scaling. This is possible in the TDS200, TDS1000, TDS2000,
TDS1000B, TDS2000B, or TDS2000C Series. The TPS2000 Series will never
return zero.
An example of unknown vertical scaling is a math waveform of CH1+CH2
where CH1 vertical scaling is different from CH2 vertical scaling. In this case,
WFMPre:YUNit? returns U. (Does not apply to the TPS2000 Series.)
WFMPre:<wfm>:YMUlt
Same as WFMPre:YMUlt, except that <wfm> specifies the source/destination
waveform instead of DATa:SOUrce and DATa:DESTINATION.
For set commands, if <wfm> is not a reference waveform, the oscilloscope
generates error 2241.
WFMPre:YOFf
YOFf is a value, expressed in digitizer levels, used to convert waveform record
values to YUNit values using the following formula (where dl is digitizer levels):
value_in_YUNits = ((curve_in_dl - YOFF_in_dl) * YMUlt) + YZERO_in_YUNits
The set form of this command stores a value for the reference waveform specified
by the DATa:DESTination command. This value does not affect how the
oscilloscope displays the waveform, but does affect the cursor readouts.
The q uery form returns a value for the waveform specified by the DATa:SOUrce
command, in digitizer levels, if that waveform is active o r displayed. If that
waveform is not active or displayed, the query fails and the oscilloscope generates
an execution error with event code 2244 (waveform requested is not active).
Group
Waveform
Syntax
WFMPre:YOFf <NR3>
WFMPre:YOFf?
Arguments
<NR3> is a value expressed in digitizing levels.
2-192 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer