WFMPre:XUNit
For all model and firmware combinations except the TDS200 Series with a
TDS2CMA communications module, the set form of this command specifies the
horizontal units ("s" for seconds and "Hz" for Hertz) for the reference waveform
specified by the DATa:DESTination command. Setting a reference waveform to
Hz causes th
e oscilloscope to display the waveform as an FFT waveform.
NOTE. For TDS200 oscilloscopes used with a TDS2CMA communications
module, the set form is ignored. Waveform commands:WFMPRE:XUNIT
WFMPRE:XUNIT
The query form returns the horizontal units for the waveform specified by
DATa:SOU
Rce. If the waveform specified by DATa:SOURce is not displayed, the
query generates event message 2244.
Group
Wavefo
rm
Syntax
WFMPre:XUNit <Qstring>
WFMPre:XUNit?
Arguments
<Qstring> is "s" or "Hz"
WFMPre:<wfm>:XUNit
Same as WFMPre:XUNit, except that <wfm> specifies the source/destination
waveform instead of DATa:SOUrce and DATa:DESTINATION.
WFMPre:XZEro
The set form of this command specifies the position, in XUNits, of the first
sample of the reference waveform spe cified by the DATa:DESTination command,
relative to the trigger.
The query form returns the position of the first sample of the waveform specified
by the DATa:SOUrce command, if that w aveform is active or displayed.
If that waveform is not active o r d isplayed, the query fails and the oscilloscope
generates an execution error with event code 2244 (waveform requested is not
active).
2-190 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer