User Manual

Command Descriptions
Examples
VERBOSE ON sets
the Verbose state true.
VERBOSE? might return the value 1, showing that the Verbose state is true.
*WAI (No Quer
yForm)
Prevents the o scilloscope from executing further commands or queries until all
pending commands that generate an OPC message are complete. This c ommand
allows you t
o synchronize the operation of the oscilloscope with your application
program. (See page 3-7, Synchronization Methods.)
The
*WAI c
ommand will stop execution until certain oscilloscope operations are
complete. (See Table 3-3 on page 3-7.)
Group
Status and Error
Syntax
*WAI
Related Commands
BUSY?, *OPC
Examples
*WAI prevents the oscilloscope from executing any further commands or queries
until all pending commands that generate an OPC message are complete.
WAVEFORMANALYSIS:SOUrce
Sets or queries the Waveform Analysis source. This command is equivalent to
setting the Sources option in the Waveform Analysis menu.
Conditions
This command applies to the TPS2000 Se ries with TPS2PWR1 Power Analysis
Module only.
Group
Power Measurement
Syntax
WAVEFORMANALYSIS:SOUrce { CH1 | CH2 | CH3 | CH4 | MATH }
WAVEFORMANALYSIS:SOUrce?
Arguments
CH1, CH2, CH3,orCH4 species one of the input channels as the waveform
analysis source.
2-180 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer