User Manual

Command Descriptions
Arguments
<NR3> the main t
rigger level, in volts.
Examples
TRIGGER:MAIN:LEVEL might return 1.4, indicating that the main edge trigger is
set to 1.4 V.
TRIGger:MAIn:MODe
Sets or queries the trigger mode for the Edge (all models) and Pulse width
(TDS1000, TDS2000, TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series
only) trigger types.
Group
Trigger
Syntax
TRIGger:MAIn:MODe { AUTO | NORMal }
TRIGger:MAIn:MODe?
Related Commands
ACQuire:STOPAfter
Arguments
AUTO generates a trigger if a trigger is not detected within a specic time period.
AUTO al so enables scan mode for sweep speeds of 100 ms/div and slower.
NORMal waits for a valid trigger event.
Examples
TRIGGER:MAIN:MODE AUTO species that a trigger event is automatically
generated.
TRIGger:MAIn:PULse? (Query Only)
Returns the current Pulse Trigger settings.
Group
Trigger
Syntax
TRIGger:MAIn:PULse?
Examples
TRIGGER:MAIN:PULSE? might return the following string:
TRIGGER:MAIN:PULSE:SOURCE CH1;WIDTH:POLARITY POSITIVE;WHEN
EQUAL;WIDTH 1.0E-3
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-171