User Manual
Command Descriptions
Group
Trigger
Syntax
TRIGger:MAIn:HOLDOff?
Returns
Main trigger holdoff value
Examples
TRIGGER:MAIN:HOLDOFF? might return TRIGGER:MAIN:HOLDOFF:VALUE
5.0E-7
TRIGger:MAIn:HOLDOff:VALue
Sets or queries the oscilloscope trigger holdoff value.
Group
Trigger
Syntax
TRIGger:MAIn:HOLDOff:VALu e <NR3>
TRIGger:MAIn:HOLDOff:VALu e?
Arguments
<NR3> is the main trigger holdoff v alue. The range is 500 ns to 10 s.
Examples
TRIGGER:MAIN:HOLDOFF:VALU E 10 sets the holdoff value to 10s.
TRIGger:MAIn:LEVel
Sets the oscilloscope edge (all models) and pulse width (TDS1000, TDS2000,
TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series only) trigger level.
This command is equivalent to adjusting the front-panel TRIGGER LEVEL knob.
NOTE. When the edge trigger source is set to AC LINE, the oscilloscope ignores
the set form of the command and generates event 221 (Settings conflict).
When the edge trigger source is set to AC LINE, the query form of the comma nd
returns zero.
Group
Trigger
Syntax
TRIGger:MAIn:LEVel <NR3>
TRIGger:MAIn:LEVel?
2-170 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer