User Manual
Command Descriptions
Syntax
TRIGger:MAIn S
ETLevel
TRIGger:MAIn?
Arguments
SETLevel sets the main trigger level to half way between the MIN and MAX
amplitudes of the trigger source input. This is equivalent to pressing the
front-panelSET LEVEL TO 50% button.
If the oscilloscope acquisition state is STOP and you send TRIGger:MAIn
SETLevel, the oscilloscope ignores the command and generates event 221
(Settings conflict).
Examples
TRIGGER:MAIN SETLEVEL sets the main trigger level mid way between MAX
and MIN.
TRIGger:MAIn:EDGE? (Query Only)
Returns the trigger coupling, source, and slope settings for the edge trigger.
Group
Trigger
Syntax
TRIGger:MAIn:EDGE?
Returns
Trigger coupling, source, and slope settings for the main edge trigger
Examples
TRIGGER:MAIN:EDGE? might return SOURCE CH1;COUPLING DC;SLOPE
RISE
TRIGger:MAIn:EDGE:COUPling
Sets or queries the type of coupling for the edge trigger. This is equivalent to
setting theCoupling optionin the Trigger menu.
Group
Trigger
Syntax
TRIGger:MAIn:EDGE:COUPling { AC | DC | HFRej | LFRej |
NOISErej }
TRIGger:MAIn:EDGE:COUPling ?
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-167