Command Descriptions
Arguments
ON enables the s
witching loss menu.
OFF disables the switching loss menu.
Examples
SWLOSS:ENABLE ON sets the switching loss measurement to on.
SWLoss:LEVELS (No Query Form)
Returns the Switching Loss measurement levels to factory default values for the
following commands:
SWLoss:TONSTART SWLoss:TONEND SWLoss:TOFFSTART
SWLoss:TOFFEND
Conditions
This command applies to the TPS2000 Series with TPS2PWR1 Power Analysis
Module only.
Group
Power Measurement
Syntax
SWLoss:LEVELS { DEFault }
Arguments
DEFault sets Switching Loss measurement levels to default values.
Examples
SWLOSS:LEVELS DEFAULT returns Switching Loss measurement levels to default
values.
SWLoss:LOG:CONDUCTION? (Query Only)
Returns the Conduction Loss for the switching loss measurement specied by
the SWLoss:LOG:INDEX command.
NOTE. The oscilloscope generates error 221 if the SWLoss:LOG:INDEX?
command returns a number greater than that reported by SWLoss:AVErage:N?
This indicates that insufcient measurements have been taken to generate a log
report.
Conditions
This command applies to the TPS2000 Series with TPS2PWR1 Power Analysis
Module only.