User Manual
Command Descriptions
SWLoss:AVEra
ge:TOTAL? (Query Only)
Returns the sum of the Turn-On, Turn-Off, and Conduction switching losses.
The oscilloscope reports this measurement in the Avg column of the Switching
Loss display
.
NOTE. This command returns error 221 (settings conflict) if the Switching Loss
menu is not active.
Conditions
This command applies to the TPS2000 Series with TPS2PWR1 Power Analysis
Module only.
Group
Power Measurement
Syntax
SWLoss:AVErage:TOTAL?
Returns
<NR3>
Examples
SWLOSS:AVERAGE:TOTAL? might return 9.1715589532E-2.
SWLoss:AVErage:TURNOFF? (Query O nly)
Returns the power loss of the device under test w hen the device is transitioning
between its on and off state. The oscilloscope reports this measurement in the Avg
column of the Switching Loss display.
You must enter the saturation voltage of the device under test using the
SWLoss:VSAT command.
NOTE. This command returns error 221 (settings conflict) if the Switching Loss
menu is not active.
Conditions
This command applies to the TPS2000 Series with TPS2PWR1 Power Analysis
Module only.
Group
Power Measurement
Syntax
SWLoss:AVErage:TURNOFF?
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-153