User Manual
Command Descriptions
Arguments
OFF or <NR1> =0d
eactivates the display of the specified waveform.
ON or <NR1> ≠ 0 a ctivates the display of the specified waveform.
<wfm> can be CH<x>, MATH, or REF<y>.
Examples
SELECT:CH2 ON displays channel 2.
SELECT:REFA? returns either 0 or 1, indicating whether the REFA waveform
is displayed.
SET? (Query O nly)
Returns most oscilloscope settings. You can send these responses back to the
oscilloscope to return the oscilloscope to the state it was in when you sent SET.
This query is identical to the *LRN? query.
NOTE. The SET? query always returns comm and headers, regardless of the
setting o f the HEADer command. This is because the returned data is intended
to be able
to be sent back to the oscilloscope as concatenated commands. The
VERBose command can still be used to specify whether the re turned headers
should be abbreviated or full length.
Group
Miscellaneous
Syntax
SET?
Related Com
mands
HEADer, *LRN?,
Returns
Most oscilloscope settings. See Appendix B: Factory Setup.
Examples
SET? might return a partial string like the following: ACQUIRE:STOPAFT ER
RUNSTOP;STAT
E 1;MOD E SAMPLE; NUMAVG 16;:HEADER
1;:VERBOSE 1;:DISPLAY:FORM AT YT;STYLE VECTORS;PERSISTENCE
500.0E-3;CONTRAST 50;:LOCK NONE;:HARDCOPY:FORMAT EPSON;PORT
RS232;LAYOUT PORTRAIT;
*SRE (No Query Form)
(Service Request Enable) sets and queries the bits in the Service Request Enable
Register (
SRER). Refer to the Status and Events chapter for more information.
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-149