User Manual

Command Descriptions
Arguments
<NR1> is an inte
ger value in the range from 1 to 5 (TDS200 Series), or 1 to 10
(TDS1000, TDS2000, TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series),
and species a memory location. Any settings that have been stored previously at
this location are overwritten.
<file p ath> (TDS2MEM, TDS1000B, TDS2000B, TDS2000C, and
TPS2000 Series only) is a quoted string that denes the path and name of the
setup le to save. Use the extension .SET for oscilloscope setup le names.
If you do not specify a path to a folder, the TDS2MEM saves the setup le in
the current working folder. The current folder refers to the name of a folder as
returned by the FILESystem:CWD query.
Examples
SAVE:SETUP 5 saves the current front-panel setup to me mory location 5.
SAVE:SETUP "A:\P ROD-TST\VID-EVA L.SET" saves the oscilloscope settings
to the le VID-EVAL.SET in the folder A:\PROD-TST on the CompactFlash card
(TDS2MEM and TPS2000 Series only), or on the USB ash drive (TDS1000B,
TDS2000B, and TDS2000C Series only).
SAVe:WAVEform (No Query Form)
Stores a waveform in one of the nonvolatile reference memory locations, or to
a le on the CompactFlash card (TDS2MEM and TPS2000 Series only), or on
the USB ash drive (TDS1000B, TDS2000B, and TDS2000C Series only). This
command is equivalent to selecting the Save Waveform option in the Save/Recall
Waveform menu.
Group
Save and Recall
Syntax
SAVe:WAVEform <wfm>, {REF<x> | <file pat h>}
Arguments
<wfm> is CH<y> (one of the allowable channels) or MATH. This
is the waveform
that will be saved. For TDS2MEM module and TPS2000 Series, <wfm> can
also be a reference waveform location (for example REF4) if the destination
is <le path>.
REF<x> is one of the allowable reference waveform storage locations.
<file p ath> (TDS2MEM, TDS1000B, TDS2000B, TDS2000C, and
TPS2000 Series only) is a quoted string that denes the path and name of the
waveform le to save on the CompactFlash card (TDS2MEM and TPS2000 Series
only), or on the USB ash drive (TDS1000B, TDS2000B, and TDS2000C Series
only). Use the extension .CSV for saved waveform les. Waveform data is saved
as self-documented comma -separated ASCII values.
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-147