User Manual
Command Descriptions
The *RST comman
d does not change the following items:
State of the RS-232, GPIB, or USB interface
Calibration data that affects device specifications
Current GPIB oscilloscope address
Stored settings or waveforms
Output queue
Service Re
quest Enable Register settings
Standard Event Status Enable Register settings
Power-On Status Clear flag setting
Hard copy parameters
Response Header Enable State
Front panel LOCK state
Verbos
e State
Display Brightness, TPS2000 Series only
Display Contrast (TDS1000, TDS2000, TDS1000B, TDS2000B, and
TPS2000 Series only)
Group
Miscellaneous
Syntax
*RST
Related Commands
FACtory, *PSC, *RCL, RECAll:SETUp, *SAV, SAVe:SETUp
*SAV (No Query Form)
Saves the state of the oscilloscope into a specified nonvolatile memory location.
You can later use the *RCL command to restore the oscilloscope to this saved
state. This is equivalent to selecting the Save Setup option in the Save/Recall
Setup menu.
Group
Save and Recall
Syntax
*SAV <N R1>
2-144 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer