Command Descriptions
the oscillosco
pe accepts all four terminators, regardless of the currently selected
terminator. When a combination of multiple characters is selected (CRLF or
LFCR), the oscilloscope interprets the first character as the terminator; it treats the
second character as a null command.
CR represents an ASCII carriage return character (0x0D) and LF represents an
ASCII linefeed character (0x0A).
Conditions
This command applies to the TDS200, TDS1000, TDS2000, and TPS2000 Series
only.
Group
RS-232
Syntax
RS232:TRANsmit:TERMinator { CR | LF | CRLf | LFCr }
RS232:TRANsmit:TERMinator?
Arguments
CR selects the carriage return character as the EOL terminator.
LF selects the line feed character as the EOL terminator.
CRLf selects the carriage return and line feed characters as the EOL terminator.
LFCr selects the line feed and carriage return c haracters as the EOL terminator.
Examples
RS232:TRANSMIT:TERMINATOR CR sets the carriage return as the EOL
terminator.
*RST
(Reset) Returns the oscilloscope to a known set of oscilloscope settings, but does
not purge any stored settings. This command executes a subset of the FACtory
command.
Sending the *RST command does the following:
Returns the oscilloscope settings to the factory defaults (refer to Appendix
B: Factory Setup)
Sets the m acro defined by *DDT to a zero-length field
Clears the pending operation flag and associated operations
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-143