User Manual

Command Descriptions
Arguments
<NR1> is an inte
ger value in the range from 1 to 5 (TDS200 Series), or 1 to 10
(TDS1000, TDS2000, TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series),
and speci es a setup storage location.
Examples
*RCL 3 restores the oscilloscope from a copy of the settings stored in memory
location 3.
RECAll:SETUp (No Query Form)
Restores a stored or factory front-panel setup of the oscilloscope from internal
nonvolatile memory, or from a le on the CompactFlas h card (TDS2MEM and
TPS2000 Series only), or on the USB ash drive (TDS1000B, TDS2000B, and
TDS2000C Series only). When used with the FACTORY argument, this command
is equivalent to pushing the DEFAULT SETUP front-panel button.
Group
Save and Recall
Syntax
RECAll:SETUp { FACtory | <NR1> | <file path> }
Related Commands
FACtory, *RCL, *RST, *SAV
Arguments
FACtory selects the factory setup.
<NR1> is a value in the range from 1 to 5 (TDS200 Series), or from 1 to 10
(TDS1000, TDS2000, TDS1000B, TDS2000B, TDS2000C, and TPS2000 Series),
and speci es a setup storage location.
<file p ath> (TDS2MEM, TDS1000B, TDS2000B, TDS2000C, and
TPS2000 Series only) is a quoted string that denes the location and name of the
setup le to recall from the CompactFlash card or the USB ash drive. If you do
not specify a path
to the folder, the TDS2MEM looks for the le in the current
working folder. The current folder refers to the name of a folder as returned by
the FILESystem:CWD query.
Examples
RECALL:SETUP FACTOR Y recalls the oscilloscope setup to its factory defaults.
RECALL:SETUP "A:\TSTSETUP\PRTST01.SET" recalls the oscilloscope
settings from the le PRTST01.SET le on the CompactFlash card (TDS2MEM
and TPS2000 Series only), or on the USB ash drive (TDS1000B, TDS2000B,
and TDS2000C Series only).
2-138 TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer