User Manual

Command Descriptions
*PSC (No Query
Form)
Sets and queries the power-on status ag that controls the automatic power-on
handling of the DESER, SRER, and ESER registers. When *PSC is true, the
DESER regist
er is set to 255 and the SRER and ESER registers are set to 0 at
power on. When *PSC is false, the current values in the DESER, SRER, and
ESER registers are preserved in nonvolatile memory when power is shut off
and are restored at power on. Refer to the Status and Events chapter for more
information.
Group
Status and Error
Syntax
*PSC <N R1>
*PSC?
Related Commands
DESE, *ESE, FACtory, *RST,
Arguments
<NR1> = 0 sets the power-on status clear ag to false, disables the power on clear,
and allows the oscilloscope to possibly assert SRQ after power on.
<NR1> 0 sets the power-on status clear ag true. Sending *PSC 1, therefore,
enables the power-on status clear and prevents any SRQ assertion after power on.
Examples
*PSC 0 sets the power-on status clear ag to false.
*PSC? might return the value 1, showing that the power-on status clear ag
is set to true.
*RCL (No Query Form)
Restores the state of the oscilloscope from a copy of its settings stored in memory.
(The settings are stored using the *SAV command.) This command is equivalent
to RECAll:SETUp, and performs the same function as theRecall Saved Setup item
in the front-panel Save/Recall Setup menu.
Group
Save and Recall
Syntax
*RCL <N R1>
Related Commands
FACtory, *LRN?, RECAll:SETUp, *RST, *SAV,
TDS200, TDS1000/2000, TDS1000B/2000B, TDS2000C, TPS2000 Programmer 2-137